Page 7 h) Install pivot mounting bracket and mount junction box on the bracket. Insert pipe/sensor assembly through the bracket to the desired po
Page 8 4.0 DESCRIPTION OF FUNCTIONS 4.1 Overview 4.1.1 The DOCN600 is a microprocessor based dissol
Page 9 Some of these are: • Recall and easy adjustment of relay and output parameters • Push button calibration •
Page 10 4.6 Analog Outputs 4.6.1 The analog output signals consist of a non-isolated 0-1 mA, 0-5 Vdc, and isolated 4-20 mA signal
Page 11 4.7.4 When in the menu mode, the display initially shows the current value of the parameters, such as the Setpoint of the control relay,
Page 12 Dip Switch Bank S2 1 Fail-safe for Relay B NO YES 2 Direction of Control for Relay B Rising Falling 3 Reserved 4 Re
Page 13 5.0 START-UP AND OPERATION 5.1 Password 5.1.1 To enter the menu press CALL then PASSWORD will be indicate
Page 14 i) If you wish to change the measuring unit to ppm, press CALL a number of times to return to “%sat/ppm” and change ppm with the down but
Page 15 5.6 Temperature 5.6.1 The temperature of the process can be read at any time by entering the menu and calling for TEMPERATURE.
Page 16 5.8.2 The relay setpoint may be at any place on the scale. To establish the setpoint proceed as follows: a) Press CALL to enter the MEN
Servicing North America:U.S.A.: One Omega Drive, P.O. Box 4047ISO 9001 Certified Stamford, CT 06907-0047TEL: (203) 359-1660 FAX: (203) 359-7700e-mail:
Page 17 5.12 Alarm High 5.12.1 The instrument is fitted with a relay, which is set to activate on both high and low alarm conditi
Page 18 6.3 Reset 6.3.1 The instrument can be reset without losing calibration or any of the stored values by pressing and re
Page 19 7.3 Temperature Output 7.3.1 The 0-5 Vdc and 0-1 mA outputs may be programmed to track the temperature of the process. Th
Page 20 b) Turn on the power. Enter the Utility Menu as described in 7.2.1. c) Press CALL to indicate CALIBRATION, which is "4-20 mA Ou
Page 21 9.0 TROUBLESHOOTING AND SERVICE 9.1 Isolate the cause 9.1.1 When a measurement problem occurs, the first step is to try
Page 22 e) Now proceed to turn each of the other S43 switches On with the remaining switches Off. The display should read within 5% of 25°C with
Page 23 d) After all the above operations are performed the FAIL LED will turn off and STATUS LED will turn green and Status “0” should be obta
Page 24 STATEMENTS OF CONFORMITY FROM THE MANUFACTURER U.S.A. WARNING: This equipment generates, uses, an
WARRANTY/DISCLAIMEROMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for aperiod of 13 months from date of
M4177/0805Where Do I Find Everything I Need for Process Measurement and Control? OMEGA…Of Course!Shop online at omega.comTEMPERATUREMUThermocouple, RT
TABLE OF CONTENTS GENERAL INFORMATION 1 SPECIFICATIONS 2 INSTALLATION 4 DESCRIPTION OF FUNCTIONS 8 START-UP AND OPERATION
Page 1 MODEL DOCN600 DISSOLVED OXYGEN ANALYZERS AND MODEL DOE-601 DISSOLVED OXYGEN SENSOR INSTRUCTION MANUAL 1.0 GENERAL INFORMAT
Page 2 2.0 SPECIFICATIONS DISPLAY: 3 1/2 digit LED, 1/2" high digits MEASURING RANGES: D.O.: 0-20.00 ppm and 0-100% sa
Page 3 MODEL DOCN600 ADDITIONAL FEATURES CONTROL RELAYS: Two relays can be independently set for operation in response to rising or falling
Page 4 3.0 INSTALLATION 3.1 Location 3.1.1 Locate the instrument within the reach of the cable provided for the DOE
Page 5 3.5 Sensor Installation, Flotation Mounting 3.5.1 Refer to diagram below and proceed as follows: a) Fasten elbow to 9
Page 6 d) The sensor may be attached to float assembly in one of two ways: 1. With sensor EXTENDED FROM BOTTOM of flotation ball: