Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
K/J/T/E-TYPE ±(0.05% rdg + 0.3°C) -50°C to 1370°C
±(0.05% rdg + 0.7°C) -50°C to -210°C
±(0.05% rdg + 0.6°F) -58°F to 2498°F
±(0.05% rdg + 1.4°F) -58°F to -346°F
N-TYPE ±(0.05% rdg + 0.7°C) -50°C to 0°C
±(0.05% rdg + 0.3°C) 0°C to 1300°C
±(0.05% rdg + 1.4°F) -58°F to 32°F
±(0.05% rdg + 0.6°F) 32°F to 2372°F
R/S-TYPE ±(0.05% rdg + 2°C) 0°C to 1767°C
±(0.05% rdg + 4°F) 32°F to 3212°F
Temperature Coefficient: 0.1 times the applicable accuracy specification
per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection: 24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Reading Rate: one time per second.
Input Connector: Accepts standard miniature thermocouple connectors (flat
blades spaced 7.9mm, center to center).